Thru-Paint Mode (E-E) eliminates the Coating and only measures Wall Thickness
The ElektroPhysik MiniTest 440 Ultrasonic Wall Thickness Gauge is a sturdy and accurate ultrasonic thickness gauge based on the ultrasonic pulse run-time method.
6 ultrasonic probes with frequencies ranging between 2 and 10 MHz in different designs offer a broad spectrum of applications and thickness measurements ranging from 0.65 mm to 500 mm. A high temperature ultrasonic probe is also available, designed for measurement of hot materials with temperatures of up to 350 C.
The device is equipped for automatic detection of the intelligent ultrasonic probes and re-calibration is not required upon switching probes. The calibration data of the sensor is saved in the probe.
Uninterrupted scanning of a part for thinning is possible at the minimum mode.
| Measuring range | Echo-Echo: 3 - 25 mm (0.12 - 0.98") Pulse Echo: With standard transducer U5.0E: 0.8 - 300 mm (0.03 - 11.8")(For ranges of other transducers see table) |
| Resolution | 0.01mm (0 99.9 mm), 0.1(> 100 mm) |
| Display | 128 x 64 pixels with backlight |
| Velocity Range | 0.0492 to .3937 in/s. 1250 to 10,000 meters/secpreset sound velocity of 9 materials |
| Measuring mode | Pulse-Echo: Standard: normal operation | Minimum: detection of thinnest point | Difference: deviation of reading to preset value | Alarm Echo-Echo: Measurements through the coating Scan: High speed scan for high temperature surfaces |
| Probe | Standard transducer U5.0E: 8mm (0.31")(For diameter of other transducers see table) |
| Operation Temperature | -20 to 50C (-4 to 122F) |
| Cable | Software and data transmission cable |
| Battery Type | Two AA batteries |
| Battery Life | more than 80 hours under continuous operation |
| Weight | approx. 6.70 oz (190 g) |
| Size | 5.1 x 2.9 x 0.94" | (130 X 73 X 24 mm) |
| Accessories Included | U5.0E 5 MHz standard transducer, Operating instructions, two 1.5V AA batteries, Plastic carrying case, Coupling liquid (200 ml), cable |
| Data storage | 10 groups with max. 500 values each |
| Data interface | USB |
| Calibration | automatic |
| Housing Protection | IP40 |
| Type | Diameter | Frequency | Measuring Range |
| U5.0E | 8 mm | 5.0 MHz | 3 25 mm / 0.12 - 0.98 " (E-E) | 0.8 -300 mm / 0.031 - 11.8" (I-E) |
| U5.0 | 8 mm | 0.31" | 5.0 MHz | 0.8 350 mm | 0.031 - 13.8" |
| U2.0 | 12 mm | 0.47" | 2.0 MHz | 2.0 500 mm | 0.079 - 19.7" |
| U7.5 | 6 mm | 0.24" | 7.5 MHz | 0.65 - 50 mm | 0.026 - 1.97" |
| U10.0 | 4 mm | 0.16" | 10.0 MHz | 0.65 20 mm | 0.026 - 0.79" |
| U5.0HT | 13 mm | 0.51" | 5.0 MHz | 3.0 200 mm | 0.12 - 7.87" |
| Type | Material Temperature | Application |
| U5.0E | -12-+60 C | measurements through the coating (E-E mode), standard transducer w ith I-E mode |
| U5.0 | -12 to 60 C | on flat surfaces or huge curvature radii |
| U2.0 | -12 to 60 C | rough surfaces such as cast iron |
| U7.5 | -12 to 60 C | thin w all thickness and small curvature radii |
| U10.0 | -12 to 60 C | small tube diameters |
| U5.0HT | -12 to 350 C | for temperatures up to 350 C |
| ElektroPhysik MiniTest 440 Ultrasonic Wall Thickness Gauge - Data Sheet |